In this example the actual non linearity is about 0 031 using mv v values and 0 032 when using calculated values which is well below the specification.
Load cell calibration mv v.
This load cell sensitivity value can be used to convert the load cell s output.
The difference voltage proportional to the load then appears on the signal outputs.
Most quality load cells will come with a factory load cell calibration sheet which specifies the actual sensitivity rating for that individual load cell.
When shopping for load cells you are typically provided with two pieces of information about the load cell excitation voltage v orvdc and output mv.
This load cell sensitivity rating is specified on a load cell data sheet often in the 1 mv v 3 mv v range.
So a 2 96 mv v load cell will provide 29 6 millivolt signal at full load when excited with 10 volts.
If a load cell has a 5 volt dc vdc excitation voltage and a 0 250 mv output the resultant ratio is 250mv 5vdc.
Supply voltage v this is the excitation voltage in volts applied to the load cell wheatstone bridge strain gauge circuit.
The millivolt output of a non amplified load cell is linearly proportional to the excitation voltage.
The cell output is rated in millivolts per volt mv v of the difference voltage at full rated mechanical load.
Divide the product from step 2 2 7 volt pounds by sensitivity in millivolts volt times the excitation voltage.
Figure 3 above is an example of a morehouse calibration shear web load cell with a non linearity specification of better than 0 05 of full scale.